Recover from complete Mass Spectometer Shutdown¶
This guide will go through the procedure for setting the ARGUS-VI set up after a complete power shutdown or reboot.
In these events all of the automatic valves will be closed, the high voltage (HV) source will be turned off, the CDD will be shutoff and all of the mass spectrometer settings will be reset.
- Manually set the values for…
Acceleration Reference [KV]
Y-Symmetry [%]
Magnet Field Reference [V]
Deflection H2 [V]
Deflection H1 [V]
Deflection AX [V]
Deflection L1 [V]
Deflection L2 [V]